Water Quality Testing · GC

Analysis solution for chloroform and carbon tetrachloride in drinking water

Drinking water samples are placed in sealed headspace vials to establish gas-liquid or gas-solid equilibrium; the gas phase in the vial is introduced into the GC to determine chloroform, carbon tetrachloride, and the volatile halogenated hydrocarbons covered by the method. This page focuses only on the two key targets, chloroform and carbon tetrachloride in drinking water, highlighting dual-target calibration, peak identification, and low-background control. Quantification and quality control use the following arrangements: calibration relationships are established using standards matched to the detection channel for chloroform, carbon tetrachloride, and the volatile halogenated hydrocarbons covered by the method; the qualitative basis, quantification mode, and result units are all based on the applicable method; blanks, calibration checks, duplicate samples, and QC samples form batch quality control. Analysis batches also include blanks, calibration checks, and applicable QC samples.

01 Solution Summary

Drinking water samples are placed in sealed headspace vials to establish gas-liquid or gas-solid equilibrium; the gas phase in the vial is introduced into the GC to determine chloroform, carbon tetrachloride, and the volatile halogenated hydrocarbons covered by the method. This page focuses only on the two key targets, chloroform and carbon tetrachloride in drinking water, highlighting dual-target calibration, peak identification, and low-background control. Quantification and quality control use the following arrangements: calibration relationships are established using standards matched to the detection channel for chloroform, carbon tetrachloride, and the volatile halogenated hydrocarbons covered by the method; the qualitative basis, quantification mode, and result units are all based on the applicable method; blanks, calibration checks, duplicate samples, and QC samples form batch quality control. Analysis batches also include blanks, calibration checks, and applicable QC samples.

IndustriesWater Quality Testing
TechnologyGC
Sample PreparationHeadspace sampling
Injection/determination interfaceQuantitative headspace gas injection
Specific SamplesDrinking water samples
AnalytesChloroform, carbon tetrachloride, and the volatile halogenated hydrocarbons covered by the method
Detection UnitECD

02 Standards and Method Basis

Method basis: application method. The laboratory may enter the name and version of the adopted formal method into the project file.

03 Samples and Analytes

The specific samples areDrinking water samples, and the analytes areChloroform, carbon tetrachloride, and the volatile halogenated hydrocarbons covered by the method. Pretreatment, injection, and detection configurations are determined according to the sample state, target concentration, and the formal method adopted.

04 Method Principle

Volatile components in a sealed headspace vial reach reproducible partitioning between the sample phase and the gas phase; the automatic headspace system samples from the gas phase and transfers it to the GC.

This route is centered on the equilibrium and gas-phase transfer of volatile components, with ECD as the detection unit.

05 Sample Collection and Pretreatment

Weigh or measure a drinking water sample into a clean headspace vial, add matrix adjustment components according to the official method used, and seal immediately. Sample vials, blank vials, and standard vials use a consistent equilibration and injection procedure.

Water samples are stored sealed and use a consistent headspace vial loading approach; the vial volume and equilibration procedure remain consistent for standard solutions, blank water, and samples.

06 Separation and Detection

The headspace gas phase enters the GC through the transfer path, the chromatographic column separates trichloromethane, carbon tetrachloride, and volatile halogenated hydrocarbons covered by the method, and the ECD acquires the response.

Chromatographic separation needs to distinguish chloroform and carbon tetrachloride from the solvent peak and other halogenated hydrocarbons; ECD is used to record the selective response of the two target compounds.

07 Instrument System Configuration

ModuleConfigurationRole in This Solution
Sample containerSealed headspace vial and sealing componentsEstablish a closed equilibrium system
Headspace samplerThermostatting, shaking, and gas-phase samplingTransfer volatile components
GC systemGCComplete the separation of volatile components
Detection UnitECDAcquire target analyte responses

08 Huishi Instruments system configuration

This application requiresGas chromatography (GC) analysis systemComplete the analysis of chloroform, carbon tetrachloride, and the volatile halogenated hydrocarbons covered by the method. When selecting the configuration, the injection interface, detector, and data system should be determined based on the sample state of drinking water samples, headspace gas-phase quantitative injection, and detection requirements.

09 Analysis workflow

1Sampling into vial
2Sealing
3Thermostatic equilibration
4Headspace gas-phase injection
5GC separation
6Detection and quantification

10 Qualitative, Quantitative, and Quality Control

Calibration relationships are established around trichloromethane, carbon tetrachloride, and volatile halogenated hydrocarbons covered by the method using reference materials matched to the detection channel; the qualitative basis, quantitation mode, and result units are subject to the applicable method; blanks, calibration checks, parallel samples, and QC samples form batch quality control. Blank headspace vials, parallel vials, calibration checks, and vial-to-vial repeatability are used to monitor the headspace process.

Check calibration, blanks, spike recovery, and repeatability for the two target compounds separately, and avoid using the response status of one target compound to represent the other.

11 Method and Configuration Selection

  • The sample must be suitable for achieving reproducible gas-phase partitioning in a sealed vial.
  • The equilibration procedure, vial volume, and sampling mode are kept consistent between calibration and samples.
  • This page is suitable for projects that focus only on chloroform and carbon tetrachloride.
  • When expanding to more halogenated hydrocarbons, a solution with broader coverage should be used.

12 Frequently Asked Questions

Why is headspace-GC used for trichloromethane, carbon tetrachloride, and volatile halogenated hydrocarbons covered by the method?

Headspace-GC matches the physicochemical properties, sample inlet, and detection targets of trichloromethane, carbon tetrachloride, and volatile halogenated hydrocarbons covered by the method; the complete route also covers the treatment, separation, detection, and data quality control of drinking water samples.

What samples is this solution applicable to?

The applicable samples are drinking water samples. When the sample source or matrix changes, the influence of co-existing components on pretreatment, separation, and detection response should be re-evaluated.

What components or indicators are included in the detection targets?

The analytes are chloroform, carbon tetrachloride, and the volatile halogenated hydrocarbons covered by the method; the main text and configuration are both bounded by this analyte scope.

How is quality controlled for an analytical batch?

For each batch of samples, set up method blanks, calibration verification, duplicate samples, or applicable QC samples, and retain records of sample processing, instrument conditions, calibration, and calculations.

How are the sample introduction and detection modules matched in this solution?

Quantitative headspace gas injection is responsible for introducing the treated drinking water sample into the analytical system, and the detection unit is configured according to the response characteristics of trichloromethane, carbon tetrachloride, and volatile halogenated hydrocarbons covered by the method; both must remain consistent with the method used.

Why is headspace vial sealing important?

The sealing state directly affects the partitioning of volatile components between the sample phase and the gas phase and the injection repeatability.

Why should the two target analytes be verified separately?

Chloroform and carbon tetrachloride differ in retention and response, so calibration, blanks, and spiked recovery need to be confirmed separately.

13 Related Solutions and Knowledge

Get Configuration and Technical Consultation

Please provide the sample information for the drinking water sample, the concentration ranges of trichloromethane, carbon tetrachloride, and the volatile halogenated hydrocarbons covered by the method, the testing batch size, the method basis, and the existing equipment. Huishi Instruments will use this to prepare a configuration list.

Chinese standard: English standard titles on this page are descriptive translations. The cited Chinese text is authoritative.

English edition updated: 2026-10-05.